beautypg.com

Utility:time, Utility:load – GW Instek PEL-2000 Series Programming User Manual User Manual

Page 54

background image

PEL-2000 Programming Manual

54

:UTILity:TIME

System Command

Description

Sets the date and time on the mainframe.

Syntax

:UTILity:TIME [aard]

Parameter

[aard]

1

2

3

“200811131300”

1
2
3

Year
Month/Day
Time (24 hours)

Example

:UTIL:TIME “200901031343”
Sets the time to 1:00 pm, January 3

rd,

2009.

Query Syntax

:UTIL:TIME?

Return Parameter [aard]

1

2

3

2008/11/13 13:00

1
2
3

Year
Month/Day
Time (24 hours)

Query Example

:UTIL:TIME?
2009/11/13/13:00

The date is November
13

th

, 2009. The time is

1:00 pm.

:UTILity:LOAD

System Command

Description

Sets the knob control style. The load module control
knobs can be set to operate independently (OLD style) to

the mainframe or with the mainframe (UPDATED).

Syntax

:UTILity:LOAD {OLD|0|Updated|1}

Parameter

OLD/0

Old

UPDATED/1

Updated

Example

:UTIL:LOAD 1

Set the knob style to
independent.

Query Syntax

:UTILity:LOAD?

Return Parameter

Knob style