beautypg.com

Current:dynamic:low:t1/t2, Current:dynamic:high:l1/l2 – GW Instek PEL-2000 Series Programming User Manual User Manual

Page 67

background image

COMMAND DETAILS

67

:CURRent:DYNamic:LOW:T1/T2

Channel Specific
Command

Description

Sets the low range timers T1 or T2 for CC dynamic
mode for the specific channel.

Syntax

:CURRent:DYNamic:LOW:T1/T2 [S/ms]

Parameter

[S/ms]

Time

T1/T2 0.1S

Sets the T1/T2 time to 0.1
seconds.

T1/T2 1

Sets T1/T2 to 1 second.

T1/T2 MIN

Sets the T1/T2 to the
minimum value.

T1/T2 MAX

Sets the T1/T2 time to the
maximum time

Example

:CURR:DYNA:LOW:T1 .1S

Sets the T1 time to 100
milliseconds for the
specific channel.

Query Syntax

: CURRent:DYNamic:LOW:T1/T2? [MIN|MAX]

Return Parameter [MAX|MIN]

Time

1 Unit=1 second

Returns T1/T2 time.

MAX/MIN

Returns the allowable
maximum and minimum.

Query Example

:CURR:DYN:LOW:T1?
2.5
:CURR:DYN:LOW:T1?
MIN
0.000025

Returns the T1 time of
2.5 seconds.
Returns the minimum T1
time allowable for the
specific channel.

:CURRent:DYNamic:HIGH:L1/L2

Channel Specific
Command

Description

Sets the high range current levels (Level 1 & 2) for
CC dynamic mode.