beautypg.com

Resistance:static:high:avalue/bvalue – GW Instek PEL-2000 Series Programming User Manual User Manual

Page 107

background image

COMMAND DETAILS

107

Example

:RES:STAT:LOW:RISE 0.1

Sets the rising slew rate
to 0.1A/uS.

Query Syntax

:RESistance:STATic:LOW:RISE/FALL? [MIN|MAX]

Return Parameter [MAX|MIN]

Slew rate

1 Unit =1 amp/uS

Returns the slew rate.

MAX, MIN

Returns the allowable
maximum and minimum.

Query Example

:RES:STAT:LOW:RISE?
MAX
0.8000

For low range CR mode,
the maximum value for
the rising slew rate is 0.8
A/uS for the specific
channel.

:RESistance:STATic:HIGH:AVALue/BVALue

Channel Specific
Command

Description

Sets the high range A/B Value for constant
resistance static mode.

Syntax

:RESistance:STATic:HIGH:AVALue/BVALue
[OHM]

Parameter

NRf+[OHM]

AVALue 10

Sets A Value to 10 ohms.
(high range only)

BVALue 20OHM

Sets B Value to 20 ohms.
(high Range only)

AVALue MIN

Sets A Value to the
minimum level for the
specific channel.

AVALue MAX

Sets A Value to the
maximum level for the
specific channel.

Example

:RES:STAT:HIGH:BVAL 10

Sets high range CR static
mode B Value to 10
ohms.