PEL-2000 Programming Manual
110
:RESistance:DYNamic:LOW:RISE/FALL
Channel Specific
Command
Description
Sets the low range rising/falling slew rate for CR
dynamic mode for the specific channel.
Syntax
:RESistance:DYNamic:LOW:RISE/FALL
Parameter
Slew Rate
RISE/FALL 0.8A/uS
Sets the rising/falling slew
rate to 0.8A/uS
RISE/FALL .1
Sets the rising/falling slew
rate to 0.1A/uS
RISE/FALL MIN
Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX
Sets to the fastest
rising/falling slew rate.
Example
:RES:DYNA:LOW:RISE 0.1
Sets the rising slew rate
to ~ 0.1A/uS.
Query Syntax
:RESistance:DYNamic:LOW:RISE/FALL? [MIN|MAX]
Return Parameter
Slew rate
1 Unit=1 amp/uS
Returns the slew rate.
MAX/MIN
Returns the allowable
maximum and minimum.
Query Example
:RES:DYN:LOW:FALL?
MIN
0.8000
For low range dynamic
CR mode, the minimum
value for the falling slew
rate is 0.8 A/uS for the
specific channel.