beautypg.com

Resistance:dynamic:low:rise/fall, Resistance:dynamic:low:t1/t2 – GW Instek PEL-2000 Series Programming User Manual User Manual

Page 110

background image

PEL-2000 Programming Manual

110

:RESistance:DYNamic:LOW:RISE/FALL

Channel Specific
Command

Description

Sets the low range rising/falling slew rate for CR
dynamic mode for the specific channel.

Syntax

:RESistance:DYNamic:LOW:RISE/FALL [A/uS]

Parameter

[A/uS]

Slew Rate

RISE/FALL 0.8A/uS

Sets the rising/falling slew
rate to 0.8A/uS

RISE/FALL .1

Sets the rising/falling slew
rate to 0.1A/uS

RISE/FALL MIN

Sets to the slowest
rising/falling slew rate.

RISE/FALL MAX

Sets to the fastest
rising/falling slew rate.

Example

:RES:DYNA:LOW:RISE 0.1

Sets the rising slew rate
to ~ 0.1A/uS.

Query Syntax

:RESistance:DYNamic:LOW:RISE/FALL? [MIN|MAX]

Return Parameter [MAX|MIN]

Slew rate

1 Unit=1 amp/uS

Returns the slew rate.

MAX/MIN

Returns the allowable
maximum and minimum.

Query Example

:RES:DYN:LOW:FALL?
MIN
0.8000

For low range dynamic
CR mode, the minimum
value for the falling slew
rate is 0.8 A/uS for the
specific channel.

:RESistance:DYNamic:LOW:T1/T2

Channel Specific
Command

Description

Sets the timers T1 or T2 for CR dynamic mode for
the specific channel in low range.

Syntax

:RESistance:DYNamic:LOW:T1/T2 [S|ms]