beautypg.com

Standard event status, Pel-2000 programming manual – GW Instek PEL-2000 Series Programming User Manual User Manual

Page 168

background image

PEL-2000 Programming Manual

168

Standard Event Status

Description

The Standard Event Status Registers indicate any
programming errors that occur. The Standard
Event Status Register group comprises of the Event
and Enable registers.

Standard Event Status Registers

8

3

4

2

2

1

1

0

128

7

64

6

32

5

16

4

Bit weight

Bit Position

DDE

QYE

OPC

0

0

0

CME

EXE

EVENT

DDE

QYE

OPC

0

0

0

CME

EXE

ENABLE

Error Bits

OPC

The operation complete bit is set when all
selected pending operations are
complete. This bit is set in response to the
*OPC command.

QYE

The Query Error bit is set is response to
an error reading the Output Queue. This
can be caused by trying to read the
Output Queue when there is no data
present.

DDE

The Device Dependent Error indicates a
memory error/lost memory or failure of
the self-test.

EXE

The Execution bit indicates an execution
error due to one of the following
Illegal command parameter
Parameter out of range
Invalid parameter
Command didn’t execute due to an
overriding operation condition.