COMMAND DETAILS
75
LOAD Subsystem
:LOAD[:STATe] ................................................................ 75
:LOAD:SHORt[:STATe] .................................................. 75
:LOAD:SHORt:KEY ......................................................... 76
:LOAD:PROTection? ...................................................... 76
:LOAD:PROTection:CLEar ............................................. 77
:LOAD:TIME? .................................................................. 77
:LOAD:DELay .................................................................. 78
:LOAD:TYPE .................................................................... 78
:LOAD[:STATe]
Channel
Specific Command
Description
This command turns the electronic load on/off for
the specific channel.
Syntax
:LOAD[:STATe] {ON|1|OFF|0}
Parameter
ON/1
Load On
OFF/0
Load Off
Example
:LOAD ON
Turns the specific
channel load on.
Query Syntax
:LOAD[:STATe]?
Return Parameter
Load module
1
Load is On
0
Load is Off
Query Example
:LOAD?
Turns the specific
channel load on.
:LOAD:SHORt[:STATe]
Channel Specific
Command
Description
This command shorts the electronic load on/off for
the specific channel.
Syntax
:LOAD:SHORt[:STATe] {ON|1|OFF|0}