beautypg.com

Configure:protection:power:level, Configure:protection:uvp:clear – GW Instek PEL-2000 Series Programming User Manual User Manual

Page 47

background image

COMMAND DETAILS

47

:CONFigure:PROTection:POWer:LEVel

Channel Specific
Command

Description

Sets the power protection level for the
current/specific channel. The level can be set to
any applicable level or to the channel
maximum/minimum.

Syntax

:CONFigure:PROTection:POWer:LEVel
[W]|MIN|MAX

Parameter

Power Protection Level

200

200Watts

200W

200Watts

MIN

Sets to the minimum level

MAX

Sets the power limit to the
maximum level

Example

:CONF:PROT:POW:LEV MAX

Sets the power limit to
102W (PEL2020)

Query Syntax

: CONFigure:PROTection:POWEr:LEVel? [MIN|MAX]

Return Parameter

Power protection level

1 unit = 1 watt

Returns the power
protection level in Watts.

MAX/MIN

Returns the allowable
maximum and minimum.

Query Example

:CONF:PROT:POW:LEV?
75

Power protection level is
at 75 watts.

:CONFigure:PROTection:UVP:CLEar

All Channel
Command

Description

Clears the under voltage power protection status.

Syntax

:CONFigure:PROTection:UVP:CLEar

Example

:CONF:PROT:UVP:CLE

Clears the under voltage
protection.