Current:static:rise/fall – GW Instek PEL-2000 Series Programming User Manual User Manual
Page 58
PEL-2000 Programming Manual
58
Query Example
:CURR:STAT:L2? 2
Returns the current
setting (2 A) for B Value.
:CURRent:STATic:RISE/FALL
Channel Specific
Command
Description
Sets the slew rate for constant current static mode.
The command is range dependant. If the current
range is Low, then the command will only apply to
the low range settings.
Syntax
:CURRent:STATic:RISE/FALL
Parameter
Slew Rate
RISE/FALL 0.078A/uS
Sets the rising/falling slew
rate to 0.078A/uS
RISE/FALL 1
Sets the rising/falling slew
rate to 1A/uS
RISE/FALL MIN
Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX
Sets to the fastest
rising/falling slew rate.
Example
:CURR:STAT:RISE .01
Sets the rising slew rate
to 0.01A/uS.
Query Syntax
: CURRent:STATic:RISE/FALL? [MIN|MAX]
Return Parameter
Slew rate
1 Unit=1 amp/uS
Returns the slew rate.
MAX/MIN
Returns the allowable
maximum and minimum.
Query Example
:CURR:STAT:RISE? MIN
0.078
the Minimum value for
the rising slew rate is
0.078 A/uS for the
specific channel.
Query Example
:CURR:STAT:RISE?
0.16800
The rising slew rate is
0.168 A/uS for the
specific channel.