COMMAND DETAILS

49

Example

:CONF:RESP 0

Response set to normal.

Query Syntax

: CONFigure:RESPonse?

Return Parameter

Response

0

Normal

1

Fast

Query Example

:CONF:RESP?
1

Response is Fast.

:CONFigure:RESEt

Channel Specific
Command

Description

Recalls the original factory default settings.

Syntax

:CONFigure:RESEt

Example

:CONF:RESE

:CONFigure:GROup:UNITs

Channel Specific
Command

Description

Sets or queries the number of single channel load
modules (PEL-2040 or PEL-2041) that can be used
in the parallel mode.

Syntax

CONFigure:GROup:UNITs |MIN|MAX

Parameter

Number of units

MIN

Sets to the minimum
number

MAX

Sets to the maximum
number

Example

CONF:GRO:UNIT 2

Sets the parallel mode to 2
units.

Query Syntax

CONFigure:GROup:UNITs? [MIN|MAX]

Return Parameter

Returns the number of
units

MAX/MIN

Returns the allowable
maximum and minimum.