6 viewing the properties of the labware, 7 setting the bottom tolerance, Viewing the properties of the labware – Eppendorf epBlue with MultiCon User Manual
Page 96: Setting the bottom tolerance
epBlue Labware Editor - Creating and managing labware
epBlue™ with MultiCon
English (EN)
96
8.2.6
Viewing the properties of the labware
Proceed as follows:
Prerequisites
• The
Labware Editor
is open.
Select the labware.
The properties of the labware are displayed in the
Properties
column.
8.2.7
Setting the bottom tolerance
Proceed as follows:
Prerequisites
• The
Labware Editor
is open.
1. Select the labware.
2. Click on the
symbol.
The
Edit Properties
window appears.
Abb. 8-6: Edit Properties window
Fig. 8-6:
Edit Properties
window
NOTICE! Material damage due to placing the pipette tip on the base of the vessel.
If the pipette tips touch the bottom of the tube, the bottom tolerance is too low. The labware
will become damaged and the dispensing volume will be distorted.
Test the application without liquid.
Check the bottom tolerance when exchanging pipette tips, plates and vessels.
If pipette tips become jammed at the base of the vessel, increase the bottom tolerance.
If deviations in the dispensing volume occur, increase the bottom tolerance.