beautypg.com

7 thermomixer, 1 checking the mixing function of the thermomixer, Thermomixer 11.7.1 – Eppendorf epBlue with MultiCon User Manual

Page 124: Checking the mixing function of the thermomixer

background image

epBlue settings - Checking components and making the basic settings
epBlue™ with MultiCon
English (EN)

124

11.7

Thermomixer

You can use this function to check the thermomixer.

Tab. 11-1: Maximum speed for labware on the thermomixer

11.7.1

Checking the mixing function of the thermomixer

Abb. 11-6: Mixer Control Thermomixer function

Fig. 11-6:

Mixer Control Thermomixer

function

NOTICE! Contamination of samples and the device due to high mixing frequency.
If the speed is set too high, liquid will spray out of the vessels. Labware may become loose
from the holder and fly about.

Observe the maximum speeds of the labware in the operating instructions.

Labware

Maximum speed in rpm

Tubes

1000

DWP 96

1200

DWP 384

1200

MTP 6

1000

MTP 96

2000

MTP 384

2000

PCR 96

2000

PCR 384

2000

Reservoir Rack

Cannot be mixed

Thermorack/Thermoblock

1000

Thermoadapter 96

1000

Thermoadapter 384

1000

Thermorack TMX

1300

PrepRack

1300