epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
173
0x2135
Location for
destination not
defined
Edit mode/parameter in
command
Transport
:
The destination location for
the labware to be transported
was not edited.
See explanation in
"Cause".
0x2136
Invalid entry for
seconds (1
...
59)
Edit mode/parameter in
command
Wait
:
A value beyond the allowed
range has been entered for the
seconds
parameter.
Enter a value between
0 and 59 seconds.
0x2137
Invalid entry for
minutes (1
...
99)
Edit mode/parameter in
command
Wait
:
A value above the allowable
maximum minutes has been
entered for the
minutes
parameter.
Enter a value between
0 and 99 minutes.
0x2138
Method without
active commands
Application only contains
passive commands (such as
wait, comment,
etc.).
Insert at least one active
command.
0x2139
Parameter conflict:
mix before
aspirating not
allowed in
multiaspirate mode
Pool/Pool One Destination:
Mix before aspirating not
allowed in multiaspirate mode.
Do not mix.
0x2152
Editing vac lid
option in vacuum
command not
possible because
no vac lid on the
worktable
To use the Vac Lid option in
the
vacuum
command the
worktable must be equipped
with a vac lid.
Editing the worktable
please place a Vac Lid on
position "GRIPPER".
0x2154
Parameter conflict:
Check levels time
may not be shorter
than remove lid
time
The
check levels
option can
only be performed when the
lid has been taken off the filter
plate before by the gripper.
Select a
check levels
time
greater than the
remove lid
time.
0x2170
Parameter conflict:
Parameter elution
from filter is only
possible when filter
plates have been
selected as source
To edit the
elution from filter
option a filter plate must have
been edited as source.
See explanations in
"Cause".
Code
Symptom/message Cause
Remedy