beautypg.com

Eppendorf epBlue with MultiCon User Manual

Page 147

background image

epBlue - Software error messages

epBlue™ with MultiCon

English (EN)

147

0x128E

Carrier: steps lost
in y

• Carrier collision.
• Carrier mechanics out of

order.

Ensure that the worktable
equipment in epBlue is
identical to the equipment
in epMotion, then restart
the application.

If the error persists, call
Eppendorf Service.

0x1294

Carrier: steps lost
in z

• Carrier collision.
• Carrier mechanics out of

order.

Ensure that the worktable
equipment in epBlue is
identical to the equipment
in epMotion, then restart
the application.

If the error persists, call
Eppendorf Service.

0x1295

Carrier:
steps lost in z
before picking up
tip

• Tip was still on pipette tool

when tool started to pick
up a new tip.

• Tip rack not placed

correctly on the worktable.

• Mechanical problems of

carrier.

Remove tips from tools.

Place tip rack correctly,
and plane, on the
worktable.

In other cases:

Call local Eppendorf
Service.

0x1296

Maximum number
of tool cycles
exceeded

See error message.

Have the tool serviced.

0x1297

Danger of collision When running the

programmed application, the
tool carrier system will touch
racks or other labware on the
worktable; e.g., during
pipetting, the optical sensor
may touch a long tube on the
adjacent position; possible
reasons:

• A low plate (microplate) is

located next to a high tube
rack.

• The 50

μL or 300

μL tip is

programmed to move
almost to the bottom of a
very long tube with
another long tube in the
adjacent position.

Program the labware on
the worktable so that high
and low labware are not
adjacent.

Program the labware so
that the 30

µL

or

300

µL

tip

does not have to move
deeply into a long vessel.

If possible: use higher
volumes in the long
vessels.

If possible: use longer tips
for the long vessels.

0x1298

Tool not calibrated Tool not calibrated.

Have the tool serviced.

0x1299

Invalid number of
samples

Value for number of samples
not permissible.

Insert an admissible value
for number of samples.

Code

Symptom/message Cause

Remedy