Option is set (see, P. 54) – Eppendorf epBlue with MultiCon User Manual
Page 54

epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
54
6.3.1.5 Changing the liquid type
Abb. 6-11: Sample transfer > Liquid types
Fig. 6-11:
Sample transfer
>
Liquid types
To change the liquid type, check the
Change parameter
checkbox.
Enter the aspiration speed of the liquid in the
Aspiration speed
field: 0.2
mm/s – 110
mm/s.
For viscous solutions and large aspiration volumes, select a low aspiration speed. The liquid aspiration
will be delayed. The liquid aspiration must be ended before the pipette tip withdraws from the liquid.
Select the low aspiration speed during phase separations or to prevent the distribution of sediments or
particles.
Speed_xl
Demineralized
water.
1
µL
3
µL
• Mixing at high dispensing
speed.
• High dispensing speed poses a
high risk of contamination for
small vessels.
• Good mixing in DWP 96 with
750
µL aspiration and 750
µL
dispensing.
Speed_xs
Demineralized
water.
1
µL
3
µL
• For slow aspiration from filter
plates.
• Very low aspiration speed.
Sediment will not be
dispersed.
Water
Demineralized
water.
1
µL
3
µL
• Recommended for most
applications.
• Technical data on the
systematic and random errors
were determined using this
liquid type.
Changes to the liquid type have effects on the accuracy of pipetting. Changing the presetting
can increase the dispensing deviations.
Liquid type
Optimized for
50 µL tips:
pipetting as of
50 µL tips:
dispensing as of
Remarks