8 magnetic separation, Magnetic separation – Eppendorf epBlue with MultiCon User Manual
Page 126

epBlue settings - Checking components and making the basic settings
epBlue™ with MultiCon
English (EN)
126
The current status will be displayed in the
State
area.
You can make settings in the
Set Temperature
area.
1. Enter the set temperature in the
New Temperature
field.
2. Enter the mixing time in the
New time
field.
3. Press the
Start
button.
The thermomixer tempers the labware to the set temperature. If the temperature is reached before the
mixing time has elapsed, the thermomixer maintains a constant temperature until the end of the test.
11.8
Magnetic separation
The
Magnetic Separation
function can be used to check the magnetic finger module.
Abb. 11-8: Magnetic Separation function
Fig. 11-8:
Magnetic Separation
function
Home
Move the magnetic finger module to the home
position.
Draw Out
Draw out the magnetic fingers.
Draw In
Draw in the magnetic fingers.