Eppendorf epBlue with MultiCon User Manual
Page 170

epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
170
0x2122
Parameter conflict:
mixing volume in
source too small for
this tool
The
Volume
and
Pipet. Tool
parameters of a mix procedure
for source vessels do not
match (
Volume
is less than the
lower limit of the tool's
volume range) (mix procedure
as defined in a command
Mix
or as part of a liquid transfer
command via parameter
Options
).
Edit
Volume
and
Pipet. Tool
so that
Volume
is within
the pipette tool's volume
range.
0x2123
Parameter conflict:
mixing volume in
source greater than
filling volume of
source tube or well
The
Volume
and
Source
parameters of a mix procedure
in the application do not
match (
Volume
is higher than
the maximum filling volume of
the source vessel) (mix
procedure as defined in a
command
Mix
or as part of a
liquid transfer command via
parameter
Options
).
Edit
Volume
and
Source
so
that
Volume
is within the
allowable filling volume of
the source vessel.
0x2124
Mixing cycles in
destination not
defined
Parameter
No. of Cycles
in a
mix procedure for destination
vessels was not edited in the
application (mix procedure as
part of a liquid transfer
command via parameter
Options
)
See explanation in
"Cause".
0x2125
Invalid entry for
mixing cycles in
destination
(1
...
99)
Entry for the parameter
No. of
Cycles
in a mix procedure for
destination vessels was higher
than the max. limit (1
up
to
99
cycles) (mix procedure as
part of a liquid transfer
command via parameter
Options
).
Enter a number between
1 and 99 for the
No. of
Cycles
parameter.
0x2126
Invalid entry for
mixing speed in
destination
(1
...
10)
Entry for the parameter
Speed
in a mix procedure for
destination vessels was higher
than the max. limit (1
up
to
10) (mix procedure as part
of a liquid transfer command
via parameter
Options
).
Enter a number between
1 and 10 for the
Speed
parameter.
Code
Symptom/message Cause
Remedy