Protein

Water with
1

%

albumin

(10

g/L),
0.01

%

Triton

X-100

5

µL

5

µL

• New pipette tips are

pre-wetted with the liquid to
be aspirated.

Prewetting

option is set (see

p. 54).

• The bending of the surface of

the liquid has a negative
impact on the free jet
capability of dispensing in cell
culture plates.

Protein C

Water with
1

%

albumin

(10

g/L),
0.01

%

Triton

X-100

5

µL

5

µL

• New pipette tips are

pre-wetted with the liquid to
be aspirated.

Prewetting

option is set (see

p. 54).

• Dispensing the liquid with a

greater distance to the
calculated plane surface of the
liquid (4 to 5 mm) than liquid
type

Protein

.

• All other data such as liquid

type

Protein

.

• Recommended for nutrient

media.

Rinse

Demineralized
water.
Water with low
surfactant
content.

1

µL

3

µL

• Low

aspiration

and

dispensing

speed, strongly delayed
blow-out.

• Other data such as liquid type

Water.

• This liquid type is suitable for

the mixing function of a
transfer command or if the

Mix

command is used.

• Can increase the risk of

contamination for small
vessels, e.g., PCR plates.

• Recommended in combination

with the

Mix

option for

reducing the residual moisture
in the pipette tip.

Liquid type

Optimized for

50 µL tips:
pipetting as of

50 µL tips:
dispensing as of

Remarks