6.3.1.3
Sample transfer
>
Mix
command
The
Mix
command is also available for mixing (see p. 63).
Abb. 6-10: Sample transfer > Mix
Fig. 6-10:
Sample transfer
>
Mix
Reuse tips
Returned pipette tips will be reused.
The pipette tips will be reused within an application for the same source vessel.
Is only executed if the
restore tips
option was selected in a previous command
and the restored tips are available.
The
Reuse tips
checkbox is only available with specific settings of the
Change
Tip
option.
Information on the aspiration and transfer of liquids can be found in the hardware operating
manual.
Mix before aspirating
Mix before each aspiration. Set the parameters following activation.
If you transfer a larger volume than the maximum volume of the pipette, the
liquid will be mixed before the first aspiration.
Mix after dispensing
Mix after each dispensing. Set the parameters following activation. Available
for
Transfer type
>
Pipette
.
If you transfer a larger volume than the maximum volume of the pipette, the
liquid will be mixed after the last aspiration.
No. of Cycles
Number of mixing cycles.
Speed
Mixing frequency. If no entries were made for
Speed
, the defined aspiration
speed of the liquid type will be used.
Volume
Mixing volume
Fixed height
Position of the pipette tip on the Z-axis. Fixed aspiration level and dispensing
level. Only for levels below the working volume.
Aspiration
Distance between the pipette tip and base of the vessel during aspiration. If
0
mm is entered, the distance will be corrected to 2
mm. The correction is
dependent on the bottom tolerance and type of labware.