epBlue - Software error messages

epBlue™ with MultiCon

English (EN)

157

0x190E

The following tool
is not selected in
the labware file
window: ...

Edit mode/procedure:
The chosen tool is not
available in the labware
collection that was selected
for your lab. Possible cause for
this error message: The
labware has been deselected
in the labware file window.

Select the respective
labware in the Labware
File window. You need to
have the appropriate user
rights. If you do not have
the necessary user rights
ask your administrator.

0x1910

The position of the
following labware is
not available on this
worktable.
The method was
written for another
workstation
configuration.

Edit mode/worktable: The
chosen labware is not
available in the labware
collection that had been
selected for your lab in the
software node

epMotion

.

Possible cause for this error
message: The labware has
been deselected in the

epMotion

node.

Set the respective labware
in the node

epMotion

.

0x1910

The method was
written for another
workstation
configuration.
The position of the
following labware is
not available on this
worktable.

The position of required
labware is not available on this
device, e.g., you have a 5075
LH and the application was
written on a 5075 VAC.

Load the concerned
application on a
compatible device, or

Modify the application
until it matches the
available device.

The position of the
following labware is
no longer allowed
yyy
in location: xxx

Edit mode/worktable: The
chosen labware may not be
placed on the selected
position anymore. Possible
cause for this error message:
The application has been
written with a former version
of the software.

Place the respective
labware on another
position.

0x1911

The following
labware has been
changed, so the
pattern no longer
fits: xxx

Because you can change the
order or contents

Tubs +

Modules(equip) + Holders

combination, the recent
pattern of a command may not
fit the new positions of the
tubes.

Either change the order or
contents of the

Tubs +

Modules(equip) + Holders

combination back to the
original or change the
pattern in the command.

Code

Symptom/message Cause

Remedy