0x2171
Parameter conflict:
Multidispense
mode is not allowed
when
elution from
filter
is selected
See explanation in the error
message.
See explanation in the
error message.
0x2172
Parameter conflict:
Transfer volume
must be set to zero
when the
elution
from filter
parameter has been
selected
Using the
elution from filter
option, the complete volume
contained in the filter plate
wells is always aspirated;
therefore, editing a volume to
be transferred is not possible.
Set the volume to zero
because the entry will not
have an effect in the
application run.
Code
Symptom/message Cause
Remedy