3 tool interlock, Tool interlock – Eppendorf epBlue with MultiCon User Manual
Page 120
epBlue settings - Checking components and making the basic settings
epBlue™ with MultiCon
English (EN)
120
11.3
Tool interlock
The
Tool interlock
function can be used to check the locking mechanism of the dispensing tools and
gripper.
Abb. 11-2: Tool interlock function
Fig. 11-2:
Tool interlock
function
Detect volumes
radio button
Activate the
Detect volumes
function. The
parameters set in the application for the
labware are used.
Use required minimum volumes
radio button
epBlue assumes that all vessels are filled with the
minimum required volume for this run. Use this
function for dry runs.
Input volumes manually
radio button
Deactivate the
Detect volumes
function. You have
to enter the run parameters manually.
Detect tips
checkbox
Activate the
Detect tips
function.
Check labware placement
checkbox
Activate the
Check labware placement
function.
Information on the optical sensor (see Optical sensor on p. 80).
1
Zero value sensor
field
0 = Dispensing tool not in home position.
1
= Dispensing tool is in home position.
2
Tool detection sensor
field
0 = Dispensing tool not available.
1
= Dispensing tool available.
3
Interlock
field
OPEN = Locking mechanism opened.
CLOSED
=
Locking mechanism closed.
4
Current Tool
field
Name of the tool used.
5
Refresh
button
Update fields 1 – 4
6
Lock
button
Locking the dispensing tool.
7
Unlock
button
Unlocking the dispensing tool.