beautypg.com

3 tool interlock, Tool interlock – Eppendorf epBlue with MultiCon User Manual

Page 120

background image

epBlue settings - Checking components and making the basic settings
epBlue™ with MultiCon
English (EN)

120

11.3

Tool interlock

The

Tool interlock

function can be used to check the locking mechanism of the dispensing tools and

gripper.

Abb. 11-2: Tool interlock function

Fig. 11-2:

Tool interlock

function

Detect volumes

radio button

Activate the

Detect volumes

function. The

parameters set in the application for the
labware are used.

Use required minimum volumes

radio button

epBlue assumes that all vessels are filled with the
minimum required volume for this run. Use this
function for dry runs.

Input volumes manually

radio button

Deactivate the

Detect volumes

function. You have

to enter the run parameters manually.

Detect tips

checkbox

Activate the

Detect tips

function.

Check labware placement

checkbox

Activate the

Check labware placement

function.

Information on the optical sensor (see Optical sensor on p. 80).

1

Zero value sensor

field

0 = Dispensing tool not in home position.
1

= Dispensing tool is in home position.

2

Tool detection sensor

field

0 = Dispensing tool not available.
1

= Dispensing tool available.

3

Interlock

field

OPEN = Locking mechanism opened.
CLOSED

=

Locking mechanism closed.

4

Current Tool

field

Name of the tool used.

5

Refresh

button

Update fields 1 – 4

6

Lock

button

Locking the dispensing tool.

7

Unlock

button

Unlocking the dispensing tool.