epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
148
0x129A
Tip too small
Reagent Transfer
: Used tip is
too small.
Use a larger tip.
0x129B
Source vessel too
small
Reagent Transfer
: Used source
vessel is too small.
Use a larger vessel.
0x12A2
Check levels in
vacuum command:
Wells in filter plate
not empty
Wells of the filter plate plug
while applying vacuum.
Follow the instructions on
the Control Panel display.
0x12A4
Labware in vacuum
manifold is not
composed of three
parts (filter plate,
frame, collection
labware)
When applying vacuum three
items must be in the vacuum
unit (filter plate above vac
frame above collection plate or
tub).
See explanation in
"Cause".
0x12A5
Well pattern of
upper and lower
rack do not fit
Vacuum unit: Well geometry
of filter plate and collection
plate must correspond to each
other (e.g., 96 well filter plate
and 96 well collection plate).
Exception: Collection plate
may be replaced by a tub.
See explanation in
"Cause".
0x12B4
Temperature unit in
location C1 not
active
In the
wait
command, the
wait
for temperature
parameter was
edited for location C1 even
though no
temperature
command for C1 had been
edited before.
Insert
temperature
command before
wait
command or delete the
wait for temperature
parameter for this location.
0x12B5
Temperature unit in
location C2 not
active
In the
wait command
, the
wait
for temperature
parameter was
edited for location C2 even
though no
temperature
command for C2 had been
edited before.
Insert
temperature
command before
wait
command or delete the
wait for temperature
parameter for this location.
0x12B6
Temperature unit in
location C3 not
active
In the
wait command
, the
wait
for temperature
parameter was
edited for location C3 even
though no
temperature
command for C3 had been
edited before.
Insert
temperature
command before
wait
command or delete the
wait for temperature
parameter for this location.
0x12B7
Temperature unit in
location C1 too hot
for labware
(temperature
command in this
method)
The temperature of location
C1 (according to a
temperature
command in this application)
is too high for the labware
placed on this location.
Select a labware item that
is more resistant to high
temperatures (e.g, made
from polypropylene or
polycarbonate).
Code
Symptom/message Cause
Remedy