Eppendorf epBlue with MultiCon User Manual
Page 146

epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
146
0x0C02
Volume too small
for this tool
Volume to be dispensed is too
small for the selected tool.
Possible causes:
• Errors in tool files.
• Errors in liquid type files.
Call local Eppendorf
Application.
0x0C08
Tool dimension
unknown
Tool dimension values
unknown. Labware outdated
or corrupt.
Make sure all labware is of
the latest version.
0x120A
Program aborted by
user
User pressed the
Abort
button
during program run.
Error message was
information for the user
that he aborted the
program.
0x1221
The hood was
opened while the
program was
stopped
See error message.
Close hood.
0x1222
Transfer allowance
was prematurely
deactivated during
program
initialization
See error message.
Start program again.
0x1281
Vacuum not
reached
The vacuum chamber or rack
is leaky or no rack is inserted.
Insert another rack.
In other cases, call local
Eppendorf Service.
0x1283
Vacuum not ready
No vacuum chamber available
or vacuum chamber not ready.
Integrate or connect
vacuum chamber, or:
Call local Eppendorf
Service.
0x128B
Carrier: steps lost
in x
• Carrier collision.
• Carrier mechanics out of
order.
Ensure that the worktable
equipment in epBlue is
identical to the equipment
in epMotion, then restart
the application.
If the error persists, call
Eppendorf Service.
0x128C
Carrier: final
position in y not
found
• Problems in carrier
movement in y-axis
(sluggish movement or no
movement at all).
• Light barrier for carrier in
y-axis defective.
Call Eppendorf Service.
Code
Symptom/message Cause
Remedy