Eppendorf epBlue with MultiCon User Manual
Page 161

epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
161
0x1A22, 0x1A32,
0x1A42, 0x1A52
No source or
destination selected
Edit mode/parameter in the
affected command:
Source or destination rack has
not been selected.
Select source or
destination, respectively.
0x1A40, 0x1A41,
0x1A50
8-channel tool
cannot be used for
this source rack
Edit mode/parameter in the
affected command:
Source rack does not fit the
8-channel tool (e.g., 24-well
plate or tube rack with
24
positions).
Choose another rack or
another tool.
0x1A61
8-channel tool
cannot be used for
this rack
Edit mode/parameter in
command
Mix
:
Rack does not fit the
8-channel tool (e.g., 24-well
plate or tube rack with
24
positions).
Choose another rack or
another tool.
0x1A62
No rack selected
Edit mode/parameter in
command
Mix
:
Rack has not been selected.
Select rack, respectively.
0x1A65
Invalid entry for
speed (0.2 ... 110)
Edit mode/parameter in
command
Mix
:
A value beyond the allowed
range has been entered for the
parameter
Speed
.
Enter a value between
0.2 and 110
mm/sec.
0x1A70
This position is
already occupied
Edit mode/pattern:
When editing the pattern you
have tried to select a certain
position that is already
occupied.
Follow the direction of the
edited pattern and move to
a different position.
0x1A76
8-channel tool
cannot be used for
this module rack
Edit mode/pattern:
Rack does not fit the
8-channel tool (e.g.,
Tubs +
Modules (equip) + Holders
combination with positions all
less than 8 in modules).
Choose another rack or
another tool.
0x1A77
No module rack or
tubes found
Edit mode/pattern:
Rack does not have any
positions (e.g.,
Tubs + Modules
(equip) + Holders
-combination
with positions all less than 1 in
Modules).
Choose another rack.
Code
Symptom/message Cause
Remedy