Eppendorf epBlue with MultiCon User Manual
Page 169

epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
169
0x211D
Mixing cycles in
source not defined
Parameter
No. of Cycles
in a
mix procedure was not edited
for the source in the
application (mix procedure as
defined in a command
Mix
or
as part of a liquid transfer
command via parameter
Options
).
See explanation in
"Cause".
0x211E
Invalid entry for
mixing cycles in
source (1
...
99)
Entry for the
No. of Cycles
parameter in a mix procedure
for source vessels was higher
than the max. limit (1
up
to
99
cycles) (mix procedure as
defined in a command
Mix
or
as part of a liquid transfer
command via parameter
Options
).
Enter a number between
1 and 99 for the
No. of
Cycles
parameter.
0x211F
Invalid entry for
mixing speed in
source (1
...
10)
Entry for the parameter
Speed
in a mix procedure for source
vessels was higher than the
max. limit (1
up
to
10)
(mix
procedure as defined in a
command
Mix
or as part of a
liquid transfer command via
parameter
Options
).
Enter a number between
1 and 10 for the
Speed
parameter.
0x2120
Mixing volume in
source not defined
Parameter
Volume
in a mix
procedure for source vessels
was not edited in the
application (mix procedure as
defined in a command
Mix
or
as part of a liquid transfer
command via parameter
Options
).
See explanation in
"Cause".
0x2121
Parameter conflict:
mixing volume in
source too large for
this tool
The
Volume
and
Pipet. Tool
parameters of a mix procedure
for source vessels are not in
agreement (
Volume
is higher
than the upper limit of the
tool's volume range) (mix
procedure as defined in a
command
Mix
or as part of a
liquid transfer command via
parameter
Options
).
Edit
Volume
and
Pipet. Tool
so that
Volume
is within
pipette tool's volume
range.
Code
Symptom/message Cause
Remedy