epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
57
6.3.3
Dilute
command
The
Dilute
command is used to create serial dilution series, e.g., for PCR standards. A defined volume is
pipetted several times from one vessel to the next.
Abb. 6-13: Principle of the Dilute command
Fig. 6-13: Principle of the
Dilute
command
The parameters of the command
Dilute
are identical to the command
Sample transfer
If the
Dilute
command is carried out within one labware the source positions and destination positions must
not overlap.
1. Fill the destination positions with a diluent. Use the
Reagent transfer
command.
2. Define the number of samples to be diluted using the
Number of samples
command.
3. In the pattern, define the number of dilution steps. The pattern is limited by a series or column.
4. Enter the options for the
Dilute
command.
5. To achieve a good mixture of sample and diluent, activate the
Mix
>
Mix after dispensing
option.
The sample and diluent will be mixed after each dispensing.
6. To mix the undiluted sample before the first aspiration, activate the
Mix before aspirating
option.
Proceed as follows to enter the pattern for the
Dilute
command:
1. Select the source position of the undiluted sample.
2. Select the destination positions of the dilution steps. The number of positions defines the number of
dilution steps.
3. Repeat the procedure for each source position.
Undiluted Sample
1. Dilution
2. Dilution
3. Dilution
Diluent
Diluent
Diluent