Eppendorf epBlue with MultiCon User Manual
Page 50
epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
50
6.3.1.2
Sample transfer
>
Options
command
Abb. 6-9: Sample transfer > Options
Fig. 6-9:
Sample transfer
>
Options
Aspirate from bottom
Aspirate liquid from the base of the vessel. When using several vessels with
different volumes, enter the lowest filling volume of these vessels (
Sample
transfer
>
Parameter
).
Elution from filter
Aspirate liquids from the supernatant in filter plates. Available if in the
Parameter
window the
Pipette
radio button is selected.
Dispense from top
During liquid dispensing, position the pipette tips 3
mm – 4
mm below the
edge of the vessel. When using several vessels with different volumes, enter
the highest filling volume of these vessels (
Sample transfer
>
Parameter
).
Change Tips
Defined time when the pipette tips are changed.
when command finished
Eject pipette tips after the end of the command. The liquid will be aspirated
with the same tip from all source labware positions.
before asp. for next
destination well
Eject the pipette tips before the liquid from the next source position is
aspirated. Liquid from the same source position will be aspirated with the same
tip. Liquid from the next source position will be aspirated using a new tip.
before each aspiration
Eject pipette tips prior to each aspiration. Liquid will always be aspirated with
a new tip. Additional volume will be discarded in the waste container along
with the pipette tip.
keep tips, do not change
tips
Do not eject pipette tips. Liquid will be aspirated with the same tip from all
source positions. The tips are also used in the next command. If the transfer
type changes in one of the following commands, or the
Mix
command is used,
the tips will be ejected.
after XX aspirations
Pipette tips will be ejected after a defined number of aspirations.
Restore tips
Pipette tips will be placed back into the position of the box they were removed
from.
Is only executed if the tips within an application are reused for the same source
vessel.
The
restore tips
checkbox is only available with specific settings of the
Change
Tip
option.