Eppendorf epBlue with MultiCon User Manual
Page 125
epBlue settings - Checking components and making the basic settings
epBlue™ with MultiCon
English (EN)
125
1. Define the speed and temperature in the
Parameter
area.
2. To check the mixing function, press the
Start
button.
The thermomixer mixes until the time is elapsed or until the test has been exited using the
Stop
button.
11.7.2
Checking the temperature control of the thermomixer
Abb. 11-7: Thermomixer temperature Control function
Fig. 11-7:
Thermomixer temperature Control
function
RPM
field
Speed of the thermomixer in rpm.
Time
field
Duration of the mixing process.
Remaining time
field
Remaining time in the mixing process.
Start
button
Stop
button
Desired Temperature
field
Set temperature of the thermomixer.
Actual Temperature
field
Actual temperature of the thermomixer.
Actual current
field
Current strength via the thermal module.
Time
field
Mixing time in min.
Remain time
field
Time until the end of the mixing process
(in min).
Temp reached after
field
Time until the set temperature is reached.
New Temperature
field
Enter a new set temperature. 4
°C–
95
°C
range.
New time
field
Enter a new mixing time. 0 s
- 99 min range.
Start
button
Start the test.
Fan on
button