Eppendorf epBlue with MultiCon User Manual
Page 163

epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
163
0x1AB1
Invalid entry for
seconds (0 ... 59)
Edit mode/parameter in
command
Vacuum
:
A value beyond the allowed
range has been entered for the
seconds
parameter.
Enter a value between
0 and 59 seconds.
0x1AB2
Invalid entry for
vacuum pressure
(1 ... 90) kPa and
(10 ... 900) mbar
resp.
Edit mode/parameter in
command
Vacuum
:
A value beyond the allowed
range has been entered for the
seconds
parameter.
Enter a value between
0 and 59 seconds.
0x1AE0
Invalid entry for
mixing speed
(0 ... 2000)
Edit mode/parameter in
command ”Thermomixer”: a
value beyond the allowed
range has been entered for the
parameter ”speed”.
Enter a value between
0 and 2000 rpm.
0x1AE1
Invalid entry for
minutes (0 ... 120)
Edit mode/parameter in
command ”Thermomixer”: a
value beyond the allowed
range has been entered for the
parameter ”minutes”.
Enter a value between
0 and 120 minutes.
0x1AE2
Invalid entry for
seconds (0 ... 59)
Edit mode/parameter in
command ”Thermomixer”: a
value beyond the allowed
range has been entered for the
parameter ”seconds”.
Enter a value between
0 and 59 seconds.
0x1AE3
Invalid entry for
temperature
(4 ... 110)
Edit mode/parameter in
command ”Thermomixer”: a
value beyond the allowed
range has been entered for the
parameter ”temperature”.
Enter a value between
4 and 110 degrees.
0x1C0B
Sample number too
large
Run mode:
The number of samples you
entered will fill more than one
rack (source or destination,
respectively) based on the
programmed pattern.
Start the application again
and enter a lower number
of samples; or:
Enter the edit mode and
program a pattern that
together with the number
of samples you want to run
will not extend beyond one
rack.
0x1C0D
You must clear old
pattern first
Press new pattern
Edit mode/pattern:
You tried to change a stored
pattern before deleting the old
pattern.
Delete the old pattern by
pressing the
New pat.
softkey.
Code
Symptom/message Cause
Remedy