epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
149
0x12B8
Temperature unit in
location C2 too hot
for labware
(temperature
command in this
method)
The temperature of location
C2 (according to a
temperature
command in this application)
is too high for the labware
placed on this location.
Select a labware item that
is more resistant to high
temperatures (e.g., made
from polypropylene or
polycarbonate).
0x12B9
Temperature unit in
location C3 too hot
for labware
(temperature
command in this
method)
The temperature of location
C3 (according to a
temperature
command in this application)
is too high for the labware
placed on this location.
Select a labware item that
is more resistant to high
temperatures (e.g., made
from polypropylene or
polycarbonate).
0x12BA
Temperature unit in
location C1 too hot
for labware
(temperature
command in
previous method)
The temperature of location
C1 (according to a
temperature
command in a previous
application with parameter
keep temperature after method
run
set active) is too high for
the labware placed on this
location.
Select a labware item that
is more resistant to high
temperatures (e.g., made
from polypropylene or
polycarbonate).
Change previous
application to not maintain
temperature after
application run.
0x12BB
Temperature unit in
location C2 too hot
for labware
(temperature
command in
previous method)
The temperature of location
C2 (according to a
temperature
command in a previous
application with parameter
keep temperature after method
run
set active) is too high for
the labware placed on this
location.
Select a labware item that
is more resistant to high
temperatures (e.g., made
from polypropylene or
polycarbonate).
Change previous
application to not maintain
temperature after
application run.
0x12BC
Temperature unit in
location C3 too hot
for labware
(temperature
command in
previous method)
The temperature of location
C3 (according to a
temperature
command in a previous
application with parameter
keep temperature after method
run
set active) is too high for
the labware placed on this
location.
Select a labware item that
is more resistant to high
temperatures (e.g., made
from polypropylene or
polycarbonate).
Change previous
application to not maintain
temperature after
application run.
0x12D0
Parameter conflict:
Elution volume too
large for this tool
Sample transfer with
elution
from filter
option:
Volume to be aspirated is too
large for the tip used.
Select a tip large enough
for picking up the liquid as
well as the additional
volume of air to be
aspirated when using this
option.
Code
Symptom/message Cause
Remedy