epBlue - Software error messages
epBlue™ with MultiCon
English (EN)

162

0x1A78

Number of tubes
not supported

Edit mode/pattern:
One or more Modules have 3,
5, 6, 7 or more than 8
positions. This is not
supported.

Choose another rack.

0x1A80

Invalid entry for
minutes (0 ... 99)

Edit mode/parameter in
command

Wait

:

A value beyond the allowed
range has been entered for the
parameter

minutes

.

Enter a value between
0 and 99 minutes.

0x1A81

Invalid entry for
seconds (0 ... 59)

Edit mode/parameter in
command

Wait

:

A value beyond the allowed
range has been entered for the
parameter

seconds

.

Enter a value between
0 and 59 seconds.

0x1A90

Selecting more
than one rack as
Source or as
Destination:
All source racks (or
all destination
racks, resp.) must
have the same well
pattern

Edit mode/parameter in
transfer command:
If more than one rack is
selected as source (or
destination) all source (or
destination), racks must have
the same well pattern (e.g.,
96-well plates).

See explanation in
"Cause".

0x1A90

Selecting more
than one rack as
Source or as
Destination:
All source racks (or
all destination
racks, resp.) must
have the same well
pattern

Edit mode/parameter in
transfer command:
The selected labware have a
different amount of wells.

Choose another rack.

0x1AA0

Invalid entry for
temperature
(0 ... 110)

Edit mode/parameter in
command

Temperature

:

A value beyond the allowed
range has been entered for the
parameter

Temperature

.

Enter a value between
0 and 110 degrees.

0x1AB0

Invalid entry for
minutes (0 ... 99)

Edit mode/parameter in
command

Vacuum

:

A value beyond the allowed
range has been entered for the
parameter

minutes

.

Enter a value between
0 and 99 minutes.

Code

Symptom/message Cause

Remedy