beautypg.com

Eppendorf epBlue with MultiCon User Manual

Page 172

background image

epBlue - Software error messages
epBlue™ with MultiCon
English (EN)

172

0x212D

Parameter conflict:
8-channel tool
cannot be used for
this source rack

Edit mode/parameter in liquid
handling command:
Source rack does not fit the
8-channel tool (e.g., 24-well
plate or tube rack with
24

positions).

Choose another rack or
another tool.

0x212E

Parameter conflict:
8-channel tool
cannot be used for
this destination
rack

Edit mode/parameter in liquid
handling command:
Destination rack does not fit
the 8-channel tool (e.g.,
24-well plate or tube rack with
24

positions).

Choose another rack or
another tool.

0x212F

Sample number too
large

The number of samples you
have entered will fill more
than one rack (source or
destination, respectively)
based on the programmed
pattern.

Program a pattern that
together with the number
of samples you want to run
will not extend beyond one
rack, or choose a lower
number of samples in the
corresponding

Number of

Samples

command.

0x2130

Parameter conflict:
tip cannot be used
for this source rack

The

Source rack

parameter in

the liquid handling command
does not match the selected
tool (e.g., 384-well plate and
TS_1000 or TM1000_8).

Choose another rack or
another tool.

0x2131

Parameter conflict:
tip cannot be used
for this destination
rack

The

Destination rack

parameter in the liquid
handling command does not
match the selected tool (e.g.,
384-well plate and TS_1000 or
TM1000_8).

Choose another rack or
another tool.

0x2132

Invalid number of
samples (1

...

384)

Number of samples you have
entered is too high.

Enter a maximum number
of samples up to

384.

0x2133

Rack or plate not
defined

Edit mode/parameter in
command

Transport

:

Labware to be transported was
not edited.

See explanation in
"Cause".

0x2134

Rack not selected in
the labware file
window

The rack you edited in a
command

Transport

(parameter

labware

) is not

selected or removed in the
labware file window and
therefore is not available for
programming.

Select the rack in the
labware file window or edit
a different rack in the
application.

Code

Symptom/message Cause

Remedy