4 pool command, 5 pool one destination command, Pool command – Eppendorf epBlue with MultiCon User Manual
Page 60: Pool one destination command
epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
60
6.3.4
Pool
command
The
Pool
command can be used to combine liquids from several source positions in one or more destination
positions.
Abb. 6-14: Principle of the Pool command
Fig. 6-14: Principle of the
Pool
command
The windows and functions of the
Pool
command are identical with the
Sample transfer
If you are using the
Multiaspirate
function, liquid will be removed from the source positions until the pipette
tip is filled. The contents will then be dispensed in the destination position. In the pattern, you define which
source positions will be pooled in the destination position.
The
Number of samples
command before the
Pool
command defines the number of source positions.
Options
>
Change Tips
is set to
before asp. for next destination well
.
6.3.5
Pool one destination
command
The
Pool one destination
command can be used to combine liquids from several source positions in one
destination position. This command is a simplified
Pool
command.
The
Number of samples
command before the
Pool one destination
command defines the number of source
positions.
The
Options
>
Change Tips
option is set to
when command finished
. The pipette tips are ejected at the end of
the command.