Eppendorf epBlue with MultiCon User Manual
Page 160
epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
160
0x1A17
Invalid entry for
speed aspiration
(0.2 ... 110)
Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
speed aspiration
parameter.
Enter a value between
0.2 and 110
mm/sec.
0x1A18
Invalid entry for
speed dispense
(0.2 ... 110)
Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter
Speed Dispense
.
Enter a value between
0.2 and 110
mm/sec.
0x1A18
Invalid entry for
speed dispense
(0.2 ... 110)
Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter
Speed Dispense
.
Enter a value between
0.2 and 110
mm/sec.
0x1A19
Invalid entry for
speed blow
(0.2 ... 110)
Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter
Speed Blow
.
Enter a value between
0.2 and 110
mm/sec.
0x1A19
Invalid entry for
speed blow
(0.2 ... 110)
Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter
Speed Blow
.
Enter a value between
0.2 and 110
mm/sec.
0x1A1A
Invalid entry for
initial stroke
(0 ... 100)
Edit mode/parameter in
transfer command: a value
beyond the allowed range has
been entered for the
parameter
initial stroke
.
Enter a value between
0 and 100%.
0x1A20
8-channel tool
cannot be used for
this source rack
Edit mode/parameter in
command
Reagent Transfer
:
Source rack does not fit the
8-channel tool (e.g., 24-well
plate or tube rack with 24
positions).
Choose another rack or
another tool.
0x1A21, 0x1A30,
0x1A31, 0x1A51
8-channel tool
cannot be used for
this destination
rack
Edit mode/parameter in the
affected command:
Destination rack does not fit
the 8-channel tool (e.g.,
24-well plate or tube rack with
24 positions).
Choose another rack or
another tool.
Code
Symptom/message Cause
Remedy