epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
68
6.3.13
Vacuum
command
The
Vacuum
command is used to create a vacuum in the vacuum unit. You can aspirate liquid in filter
plates.
Prior to executing the application, equip the vacuum unit as described in the epMotion hardware operating
manual.
Abb. 6-22: Vacuum window
Fig. 6-22:
Vacuum
window
The command
Vacuum
only is available for devices with vacuum unit.
Frame
drop-down list
Type of Vac Frame used.
Pressure
field
Vacuum in the vacuum unit. Unit mbar or kPa.
Maximum underpressure 850
mbar/85
kPa.
Duration
field
Time during which the vacuum is maintained.
0
s – 100
min
59
s
range.
Use Vacuum Lid
checkbox
Activate transport of the Vac Lid with the gripper.
Use Vacuum Lid
drop-down list
Vac Lid used.
Check levels
checkbox
The optical sensor checks the remaining volume
of all wells in a filter plate on the vacuum unit. If
remaining liquid has been detected, a vacuum
will be reapplied. Checking the remaining volume
takes too long.