0x2026
Bottom tolerance
too small
Bottom tolerance too small.
Use a bigger value.
0x2101
Tool not defined
Parameter
Pipet. Tool
was not
edited in the application.
See explanation in
"Cause".
0x2104
Tips not edited in
worktable/
procedure
Tips were edited in the
procedure of the application,
but they were not edited in the
worktable (e.g., filter tips <->
tips without filter).
Edit the tips that you
programmed in the
procedure in the
worktable.
0x2105
Parameter conflict:
Start volume
greater than filling
volume of source
tube or well
The
Volume
and
Source
parameters of the source
vessel do not match (
Volume
is
higher than the maximum
filling volume of the source
vessel).
Edit
Volume
and
Source
in
the worktable so that
Volume
is covered by the
maximum filling volume of
the source vessel.
0x2106
Parameter conflict:
Start volume
greater than filling
volume of
destination tube or
well
The
Volume
and
Destination
parameters of the destination
vessel do not match (
Volume
is
higher than the maximum
filling volume of the
destination vessel).
Edit
Volume
and
Destination
in the
worktable so that
Volume
is
covered by the maximum
filling volume of the
destination vessel.
0x2107
Volume not defined Parameter
Volume
was not
edited in the application.
See explanation in
"Cause".
0x2108
Parameter conflict:
Volume too small
for this tool
The
Volume
and
Pipet. Tool
parameters of the application
do not match (
Volume
is
smaller than the lower limit of
the tool volume range).
Edit
Volume
and
Pipet. Tool
so that
Volume
is covered
by the volume range of the
pipette tool.
0x2109
Parameter conflict:
Volume too large
for this tool
The
Volume
and
Pipet. Tool
parameters of the application
do not match (
Volume
is
higher than the upper limit of
the tool volume range).
Edit
Volume
and
Pipet. Tool
so that
Volume
is covered
by the volume range of the
pipette tool.
0x210A
Parameter conflict:
volume greater
than filling volume
of source tube or
well
The
Volume
and
Source
parameters of the application
do not match (
Volume
is
higher than the maximum
filling volume of the source
vessel).
Edit
Volume
and
Source
so
that
Volume
is covered by
the maximum filling
volume of the source
vessel.
0x210B
Parameter conflict:
volume greater
than filling volume
of destination tube
or well
The
Volume
and
Destination
parameters of the application
do not match (
Volume
is
higher than the maximum
filling volume of the
destination vessel).
Edit
Volume
and
Destination
so that
Volume
is covered by the
maximum filling volume of
the destination vessel.
Code