4.2.7
Entering the filling level of the labware
4.2.7.1 Checking the filling level automatically
If the
Detect volumes
function is active, the optical sensor checks the filling level after the start.
If the following problems occur, a dialog window appears:
• Filling level too low.
• Filling level too high.
• Filling level cannot be correctly detected.
Abb. 4-7: Checking window
Fig. 4-7:
Checking
window
If the
Calculated volume
is larger than the
Maximum volume
, remove liquid from the vessel.
If the
Calculated volume
is smaller than the
Minimum volume
, add liquid to the vessel.
The optical sensor cannot carry out a detection if there is an air bubble in the liquid. Remove the air
bubble. Select
Repeat scan
to repeat the detection.
Maximum volume
Filling level of the vessel.
Minimum volume
Volume required to remove the set liquid
quantity.
Calculated volume
Volume in the vessel, calculated using vessel
data and the liquid level.
Repeat scan
button
Repeat
Detect volumes
.
Abort
button
Aborts the application.
User input
button
Manual volume entry.
Accept level and continue
button
Accept the displayed calculated volume.
µL
input field
Manually select the volume if the
User Input
button has been selected.
NOTICE! Dispensing error due to refilling of vessels.
The optical sensor only checks the level of the vessels at the specified location.
Only change the volume of the vessels at the specified location.
Do not change the volume of the vessels at other locations.