epBlue™ with MultiCon

English (EN)

31

4.2.7

Entering the filling level of the labware

4.2.7.1 Checking the filling level automatically

If the

Detect volumes

function is active, the optical sensor checks the filling level after the start.

If the following problems occur, a dialog window appears:

• Filling level too low.
• Filling level too high.
• Filling level cannot be correctly detected.

Abb. 4-7: Checking window

Fig. 4-7:

Checking

window

If the

Calculated volume

is larger than the

Maximum volume

, remove liquid from the vessel.

If the

Calculated volume

is smaller than the

Minimum volume

, add liquid to the vessel.

The optical sensor cannot carry out a detection if there is an air bubble in the liquid. Remove the air
bubble. Select

Repeat scan

to repeat the detection.

Maximum volume

Filling level of the vessel.

Minimum volume

Volume required to remove the set liquid
quantity.

Calculated volume

Volume in the vessel, calculated using vessel
data and the liquid level.

Repeat scan

button

Repeat

Detect volumes

.

Abort

button

Aborts the application.

User input

button

Manual volume entry.

Accept level and continue

button

Accept the displayed calculated volume.

µL

input field

Manually select the volume if the

User Input

button has been selected.

NOTICE! Dispensing error due to refilling of vessels.
The optical sensor only checks the level of the vessels at the specified location.

Only change the volume of the vessels at the specified location.

Do not change the volume of the vessels at other locations.