epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
49
6.3.1.1
Sample transfer
>
Parameter
command
Abb. 6-8: Sample transfer > Parameter
Fig. 6-8:
Sample transfer
>
Parameter
Volume
Select volume in µL or nL. A decimal place is available for volumes ≤
99.9
µL.
You can transfer a volume larger than the maximum volume of the pipette tip.
The volume to be transferred will be pipetted in several steps.
Example: The user would like to transfer 1500
µL
and
enters
this
value
into
the
Volume
field. If a 1000
µL tip is being used, 2x
750
µL will be pipetted. If a
300
µL tip is being used, 5
x
300
µL will be pipetted.
Pipet. tool
Used dispensing tool.
Pipette
The volume is individually aspirated and dispensed for each liquid transfer.
Multidispense
Aspirate liquid and dispense it in the destination position in defined subsets.
Multiaspirate
Aspirate liquid from several source positions using one pipette tip and
dispense the liquid in one destination position. Available for
Pool
and
Pool one
destination
commands.
Use filter tips
Activate when using tips with filter.
Source
Select the source labware. The source labware must be placed on the
worktable. You can select 4 source labwares. If you are using more than one
source labware, the number of positions in the labware (384, 96, 24) must
match. If you dispense within one labware, the source labware and destination
labware are identical.
Destination
Select the destination labware. The destination must be placed on the
worktable. You can select 4 destination labwares. If you are using more than
one destination labware, the number of positions in the labware (384, 96, 24)
must match. If you dispense within one labware, the source labware and
destination labware are identical.
Pattern
Select the source positions and destination positions. Create replicates.
Information on the aspiration and transfer of liquids by means of the epMotion can be found in
the hardware operating manual.