Eppendorf epBlue with MultiCon User Manual
Page 164

epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
164
0x1C0D
You must clear old
pattern first
Press "new pattern"
Edit mode/pattern:
You tried to change a stored
pattern before deleting the old
pattern.
Delete the old pattern by
pressing the button
new
pattern
.
0x1C0E
You must go
forward
Edit mode/pattern:
When entering the pattern, the
order of edited locations in the
source (or destination,
respectively) must be from left
to right or from the top of the
pattern downwards (i.e., move
only in columns or in rows).
See explanation in
"Cause".
0x1C0F
You may only move
horizontally or
vertically
Edit mode/pattern:
When entering the pattern the
order of edited locations in the
source (or destination, resp.)
must be from the left to right
or from top of the pattern
downwards (i.e., move only in
columns or in rows).
Note: Error message may also
occur when working with an
8-channel tool and editing
another position than the
upper ones (see error code
0x1C1F).
See explanation in
"Cause".
0x1C10
Pattern for
replicates of first
sample too complex
Edit mode/pattern:
The pattern algorithm cannot
handle this pattern.
Enter a simpler pattern if
possible.
In case this is not possible:
Call local Eppendorf
Application Support.
0x1C11
Pattern too
complex
Edit mode/pattern:
The pattern algorithm cannot
handle this pattern.
Note: See note in error
0x1C0F.
Enter a simpler pattern if
possible.
In case this is not possible:
Call local Eppendorf
Application Support.
0x1C14
Pattern must fit in
rows or columns
Edit mode/pattern:
The basic unit of the pattern
you tried to enter extends
beyond a row or a column.
This cannot be handled by the
pattern algorithm.
Enter a simpler pattern if
possible.
In case this is not possible:
Call local Eppendorf
Application Support.
Code
Symptom/message Cause
Remedy