epBlue - Software error messages

epBlue™ with MultiCon

English (EN)

165

0x1C15

Pattern too
complex

Edit mode/pattern:
The pattern algorithm cannot
handle this pattern.
Note: See note in error
0x1C0F.

Enter a simpler pattern if
possible.

In case this is not possible:

Call local Eppendorf
Service.

0x1C16

This position is
already occupied

Edit mode/pattern:
When editing the pattern you
have tried to select a certain
rack position that is already
occupied.

Following the edited
pattern move to a different
rack position.

0x1C17

You must start with
the source

Edit mode/pattern:
When editing a pattern you
must start with the source.

See explanation in
"Cause".

0x1C18

Please enter a
source now

Edit mode/pattern:
In the destination rack, you
tried to enter more replicates
than you had sources.

Enter the same number of
replicates for all sources
you edit when
programming a pattern.

0x1C19

Please enter a
destination now

Edit mode/pattern:
When having selected a
source in the

Sample Transfer

command you first have to
enter a destination for this
source before moving to the
next source position.

Press the

Destination

softkey and edit the
destination position(s) for
the selected source
position.

0x1C1A

No more positions
available (limited
by Number of
Samples
command)

Edit mode/pattern:
Editing further positions is not
possible because the limit set
in the

Number of Samples

command would be exceeded.

Select a pattern that fits the
programmed

Number of

Samples

command.

0x1C1E

Pattern for reagent
transfer:
Source can be
chosen only once

Edit mode/pattern for
command

Reagent Transfer

:

After having entered the
source and the destinations
for the reagent transfer you
cannot select an additional
source.

Enter the source only once.
In case this does not meet
your requirements for this
application consider
selecting command

Sample

Transfer

instead of

Reagent

Transfer

, or:

Call local Eppendorf
Application Support.

0x1C1F

Pattern with
8-channel tool:
Please edit upper
position of this tool

Edit mode/pattern with
8-channel tool:
Only the upper positions of the
8-channel tools can be
selected.

See explanation in
"Cause".

Code

Symptom/message Cause

Remedy