epBlue - Software error messages
epBlue™ with MultiCon
English (EN)

158

0x1911

The following
labware has been
changed, so the
pattern no longer
fits: xxx

Because the order or contents
of a “Equipped Holder”
combination can be changed,
the recent pattern of a
command may not fit the new
positions of the tubes.

Either change the order or
contents of the “Equipped
Holder” combination back
to the original or change
the pattern in the
command.

0x1A02 The

name

is

already

used for another
labware

Edit mode/labware:
The same name has been
defined for a different rack or
another labware item.

Enter a different name.

0x1A03

This position is not
available for the
selected labware

Edit mode/worktable:
Certain worktable positions
are not allowed for specific
labware (e.g., tips can only be
placed in the rear of the
worktable).

Place the selected labware
in another location.

0x1A04

The selected
labware may not be
stacked on top of
labware already
placed

Edit mode/worktable:
Building of labware stacks on
the worktable is restricted to
certain labware combinations
(e.g., thermorack above
thermorack does not make
sense).

See "Cause".

0x1A05

Maximum number
of labware stacked:
5

Edit mode/worktable:
Certain labware build a pile
which contains too many
when placed on certain
labware.

Place the selected labware in
another location.

0x1A06

Labware stack too
high in location:
xxx
Maximum height:
xxx mm

Edit mode/worktable:
Labware stacks on the
worktable may not exceed a
maximum height limit (e.g.,
plates on adapters is allowed;
reservoir holder on adapters is
not allowed because the stack
would be too high).

See "Cause".

0x1A10

8-channel tool
cannot be used for
this source rack

Edit mode/parameter in
command

Sample Transfer

:

Source rack does not fit the
8-channel tool (e.g., 24-well
plate or tube rack with
24

positions).

Choose another rack or
another tool.

Code

Symptom/message Cause

Remedy