beautypg.com

Eppendorf epBlue with MultiCon User Manual

Page 52

background image

epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)

52

6.3.1.4

Sample transfer

>

Liquid types

Liquid types are defined in epBlue.

Tab. 6-1:

Defined liquid types in epBlue

Dispense

Distance between the pipette tip and base of the vessel during dispensing. If
the dispensing position is above the vessel, the distance will be corrected to
the height of the vessel.

Information on mixing liquids can be found in the hardware operating manual.

NOTICE! Contamination of the dispensing tool due to high mixing speed.
If the mixing speed is too high, liquid may enter the dispensing tool.

If the liquid has a low viscosity or tends to foam, set a low mixing speed.

Use demineralized water to test the mixing settings.

Use pipette tips with filter.

Information on dispensing liquids can be found in the hardware operating manual.

Liquid type

Optimized for

50 µL tips:
pipetting as of

50 µL tips:
dispensing as of

Remarks

Alcohol

75%

Mixture of
75

%

ethanol

and 25

%

water

1 to 3

µL

3

µL

• Wash reagent in kits for

nucleic acid purification. See
applications.

Alcohol

98%

Alcohol 98

%

1

µL

3

µL

• New pipette tips are

pre-wetted with the liquid to
be aspirated.

Prewetting

option is set (see

p. 54).

• For multidispensing with

300

µL pipette tips, the

distance to the filter with
300

µL aspiration is very low.

To prevent the filter from
becoming wetted, pipetting
should be avoided with an
aspiration of 280

µL or more.

Glycerol

Mixture of
40

%

glycerol

and 60

%

water

1

µL

5

µL

• Liquid type for enzyme

solutions.

• If the glycerol content of the

enzyme solutions is
significantly lower, liquid type

Water

can be used.