Eppendorf epBlue with MultiCon User Manual
Page 159

epBlue - Software error messages
epBlue™ with MultiCon
English (EN)
159
0x1A10 8-channel
tool
cannot be used for
this source rack
Edit mode/parameter in
command
Sample Transfer
:
Source rack does not fit to
8-channel-tool (e.g.:
24-well-plate or tube rack with
24 positions).
Choose another rack or
another tool.
0x1A11
8-channel tool
cannot be used for
this destination
rack.
Edit mode/parameter in
command
Sample Transfer
:
Destination rack does not fit
the 8-channel tool (e.g.,
24-well plate or tube rack with
24
positions).
Choose another rack or
another tool.
0x1A12
No source or
destination selected
Edit mode/parameter in
command
Sample Transfer
:
Source or destination rack has
not been selected.
Select source or
destination, respectively.
0x1A15
Invalid entry for
movement blow
(0 ... 100)
Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter
Movement Blow
.
Enter a value between
0 and 100%.
0x1A15
Invalid entry for
movement blow
(0 ... 100)
Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter
Movement Blow
.
Enter a value between
0 and 100%.
0x1A16
Invalid entry for
delay blow
(0 ... 9999)
Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter
Delay Blow
.
Enter a value between
0 and 9999
msec.
0x1A16
Invalid entry for
delay blow
(0 ... 9999)
Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
Delay Blow
parameter.
Enter a value between
0 and 9999
msec.
0x1A17
Invalid entry for
speed aspiration
(0.2 ... 110)
Edit mode/parameter in
transfer command:
A value beyond the allowed
range has been entered for the
parameter
Speed Aspiration
.
Enter a value between
0.2 and 110
mm/sec.
Code
Symptom/message Cause
Remedy