If the

Magnetic separation

command is executed when the

Separation On

checkbox is checked, the

magnetic fingers will be drawn out. During the waiting period, the magnetic beads will be separated by the
drawn out magnetic fingers on the tube inner wall. After the waiting period has elapsed, the next command
in the procedure will be executed. If the

Magnetic separation

command is executed when the

Separation On

checkbox is not checked, the magnetic fingers will be drawn in.

Abb. 6-17: Magnetic separation command

Fig. 6-17:

Magnetic separation

command

Starting the separation

To start the separation, execute the

Magnetic separation

command when the

Separation On

checkbox is

activated.

The magnetic fingers will be drawn out.

Exiting the separation

To end the separation, execute the

Magnetic separation

command when the

Separation On

checkbox is

deactivated.

The magnetic fingers will be drawn in.

6.3.9

Mix

command

The

Mix

command can be used to mix liquids in a position by multiple aspiration and dispensing steps.

The options of the

Mix

command are identical with the options of the

Sample transfer

command in the

Mix

area (see p. 51).

Separation enabled

checkbox

Check the checkbox to start the separation.
Uncheck the checkbox to end the separation.

Duration

field

Waiting period. The application stops until the
waiting period has elapsed.