9 mix command, Mix command – Eppendorf epBlue with MultiCon User Manual
Page 63
epBlue Application Editor - Creating and editing applications
epBlue™ with MultiCon
English (EN)
63
If the
Magnetic separation
command is executed when the
Separation On
checkbox is checked, the
magnetic fingers will be drawn out. During the waiting period, the magnetic beads will be separated by the
drawn out magnetic fingers on the tube inner wall. After the waiting period has elapsed, the next command
in the procedure will be executed. If the
Magnetic separation
command is executed when the
Separation On
checkbox is not checked, the magnetic fingers will be drawn in.
Abb. 6-17: Magnetic separation command
Fig. 6-17:
Magnetic separation
command
Starting the separation
To start the separation, execute the
Magnetic separation
command when the
Separation On
checkbox is
activated.
The magnetic fingers will be drawn out.
Exiting the separation
To end the separation, execute the
Magnetic separation
command when the
Separation On
checkbox is
deactivated.
The magnetic fingers will be drawn in.
6.3.9
Mix
command
The
Mix
command can be used to mix liquids in a position by multiple aspiration and dispensing steps.
The options of the
Mix
command are identical with the options of the
Sample transfer
command in the
Mix
Separation enabled
checkbox
Check the checkbox to start the separation.
Uncheck the checkbox to end the separation.
Duration
field
Waiting period. The application stops until the
waiting period has elapsed.