beautypg.com

Inputs, Calibration offset, Calibration – Watlow EZ-ZONE RMC User Manual

Page 158

background image

Watlow EZ-ZONE

®

RMC Module

155

Chapter 7 Features

ç

WARNING! During autotuning, the controller sets the out-

put to 100 percent and attempts to drive the Process Value

toward the set point . Enter a set point and heat and cool

power limits that are within the safe operating limits of your

system .

Inputs

Calibration Offset

Calibration offset allows a user to compensate for an

inaccurate sensor, lead resistance or other factors

that affect the input value. A positive offset increases

the input value, and a negative offset decreases the

input value.

The input offset value can be viewed or changed

with Calibration Offset [`i;CA] (Operations Page,

Analog Input Menu).

Time

Temperature

Temperature Reading

from Sensor

Actual Process Temperature

Negative Calibration Offset will

compensate for the difference

between the Sensor Reading and

the Actual Temperature

Calibration

Before performing any calibration procedure, verify

that the displayed readings are not within published

specifications by inputting a known value from a pre-

cision source to the analog input. Next, subtract the

displayed value with the known value and compare

this difference to the published accuracy range speci-

fication for that type of input.

Use of the Calibration Offset [`i;CA] parameter

found in the Operations Page [oPEr] , Analog Input

Menu [``Ai] shifts the readings across the entire

displayed range by the offset value. Use this param-

eter to compensate for sensor error or sensor place-

ment error. Typically this value is set to zero.
Equipment required while performing calibra-

tion: Obtain a precision source for millivolts, volts,

milliamperes or resistance depending on the sensor

type to be calibrated. Use copper wire only to connect

the precision source to the controller’s input. Keep

leads between the precision source and controller as

short as possible to minimize error. In addition, a

precision volt/ohm meter capable of reading values to

4 decimal places or better is recommended. Prior to

calibration, connect this volt/ohm meter to the preci-

sion source to verify accuracy.

Actual input values do NOT have to be exactly the

recommended values, but it IS critical that the actual

value of the signal connected to the controller be ac-

curately known to at least four digits.
Calibration of Analog Inputs:
To calibrate an analog input, you will need to provide

a source of two electrical signals or resistance values

near the extremes of the range that the application is

likely to utilize. See recommended values below:

Sensor Type

Precision

Source Low

Precision

Source High

thermocouple

0.000 mV

50.000 mV

millivolts

0.000 mV

50.000 mV

volts

0.000V

10.000V

milliamps

0.000 mA

20.000 mA

100 Ω RTD

50.00 Ω

350.0 Ω

1,000 Ω RTD

500.0 Ω

3,500 Ω

thermistor 5 kΩ

50.00

5,000

thermistor 10 kΩ 150.0

10,000

thermistor 20 kΩ 1,800

20,000

thermistor 40 kΩ 1,700

40,000

potentiometer

0.000

1,200

Note:

The user may only calibrate one sensor type. If the

calibrator interferences with open thermocouple detec-

tion, set Sensor Type [`SEn] in Setup Page [`SEt],

Analog Input Menu [``Ai] to millivolt [`Mu] in-

stead of Thermocouple [``tC] to avoid interference

between the calibrator and open thermocouple detect

circuit for the duration of the calibration process. Be

sure to set sensor type back to the thermocouple type

utilized.

1. Disconnect the sensor from the controller.
2. Record the Calibration Offset [`i;CA] parameter

value in the Operations Page [oPEr] , Analog In-

put Menu [``Ai] then set value to zero.
3. Wire the precision source to the appropriate con-

troller input terminals to be calibrated. Do not

have any other wires connected to the input ter-

minals. Please refer to the Install and Wiring sec-

tion of this manual for the appropriate connec-

tions.
4. Ensure the controller sensor type is programmed

to the appropriate Sensor Type [`SEn] to be uti-

lized in the Setup Page [`SEt] , Analog Input

Menu

[``Ai]

.

5. Enter Factory Page [FCty] , Calibration Menu

[`CAL] via RUI or EZ-ZONE Configurator Soft-

ware.
6. Select the Calibration [CAL] input instance to be

calibrated. This corresponds to the analog input to

be calibrated.
7. Set Electrical Input Slope [ELi;S] to 1.000 and

Electrical Input Offset [ELi;o] to 0.000 (this will

cancel any prior user calibration values)