beautypg.com

Selecting waveforms (vcv only), V (figure 8-18), S (figure 8-18) are grayed out – Philips V200 User Manual

Page 97: Operating instructions

background image

Chapter 8

Operating Instructions

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

8-19

Figure 8-18: Setting Alarm Limits That Are Not Currently Active

Selecting Waveforms (VCV only)

Two inspiratory flow waveforms for mandatory VCV breaths are available:
descending ramp and square wave (Figure 8-19). The selected waveform is
highlighted and defines the inspiratory flow for all mandatory VCV breaths,
whether they are initiated by the patient, the ventilator, or the operator. The
waveform selection is not applicable to PCV or NPPV.

NOTE:

Any of the changes made in the screen shown in Figure 8-18, do
not take effect until the operator switches to the new ventilation
breath type (in this case Pressure Control).

Grayed out buttons indicate that the alarm limits are not active

Push button of
alarm setting to
adjust

Press this ACTIVATE
button to switch to the
ventilation breath type
indicated. The
ventilation breath type
on this button
indicates which alarm
set is shown on this