Options and accessories – Philips V200 User Manual
Page 225

Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-61
Chapter 13
Static C and R
Measurement out of range. Check tubing 
system for leaks 
Static compliance exceeds 0 to 350 mL/
cmH
2
O or static resistance exceeds 0 to
400 cmH
2
O/L/s:
Static C & R unavailable due to alarm 
condition 
Alarm condition exists when maneuver 
requested.
Static C & R only available in VCV 
Ventilator not in VCV mode. 
Static C & R unavailable; I-Time less than 
200ms
Inspiration time must be between 200 ms 
and 5 seconds. 
Static C & R unavailable; I-Time greater 
than 5 seconds
Inspiration time must be between 200 ms 
and 5 seconds.
Plateau pressure stability not achieved; 
Static C & R unavailable 
Stable plateau pressure cannot be 
achieved.
Maneuver cancelled 
Operator presses Stop button. 
Maneuver timed out
No machine or operator initiated inspiration 
within 20 seconds of starting maneuver.
Maneuver cancelled by alarm condition
Alarm condition occurs after maneuver 
begins.
None: all alarms are active during 
the Static C and R maneuver.
Maneuver
Error Messages
Temporarily Disabled Alarms
Table 13-8: Respiratory Mechanics Error Messages and Alarms
