Options and accessories – Philips V200 User Manual
Page 225
Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-61
Chapter 13
Static C and R
Measurement out of range. Check tubing
system for leaks
Static compliance exceeds 0 to 350 mL/
cmH
2
O or static resistance exceeds 0 to
400 cmH
2
O/L/s:
Static C & R unavailable due to alarm
condition
Alarm condition exists when maneuver
requested.
Static C & R only available in VCV
Ventilator not in VCV mode.
Static C & R unavailable; I-Time less than
200ms
Inspiration time must be between 200 ms
and 5 seconds.
Static C & R unavailable; I-Time greater
than 5 seconds
Inspiration time must be between 200 ms
and 5 seconds.
Plateau pressure stability not achieved;
Static C & R unavailable
Stable plateau pressure cannot be
achieved.
Maneuver cancelled
Operator presses Stop button.
Maneuver timed out
No machine or operator initiated inspiration
within 20 seconds of starting maneuver.
Maneuver cancelled by alarm condition
Alarm condition occurs after maneuver
begins.
None: all alarms are active during
the Static C and R maneuver.
Maneuver
Error Messages
Temporarily Disabled Alarms
Table 13-8: Respiratory Mechanics Error Messages and Alarms