Options and Accessories

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

13-91

Chapter 13

Turn the knob to highlight EXIT and then press the knob. Turn the
knob to highlight EXIT again and then press the knob.The monitor
returns to the SETUP screen.

The ventilator and the respiratory profile monitor will begin communication
within 60 seconds.

RS-232
Communications

The bidirectional communication between the respiratory profile monitor and
ventilator is initiated by the monitor. The ventilator recognizes the connect
request message from the monitor and responds.

Information Transferred from the Respiratory Profile Monitor to the
Ventilator

The respiratory profile monitor provides the ventilator with the data listed in
the following table. The monitor sends the average value for each parameter for
the previous 60 seconds.

Table 13-18: Information Transferred from the Monitor to the Ventilator

Information Transferred from the Respiratory Profile Monitor to the Ventilator

Data

Description

CO-a

Cardiac Output Average

NOTE: Available only when the Cardiac Output option is
installed.

SpO

2

O2 Saturation

ETCO

2

End Tidal CO

2

VCO

2

Volume CO

2

Vtalv Alveolar

Tidal

Volume

MValv

Alveolar Minute Volume