Chapter 13
13-104
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Options and Accessories
•
Do not use with neonatal tracheostomy tubes.
•
Do not use on neonates.
•
Do not inflate the tracheostomy tube cuff until Speaking Mode has
been turned off.
•
Do not use on patients who are dependent on PEEP therapy.
•
Do not use with HME filters.
•
Discontinue using Speaking Mode promptly and institute appropriate
ventilation if the patient experiences difficulty or their status
deteriorates from baseline parameters.
•
Perform an airway patency test immediately after activating Speaking
Mode.
NOTE:
Speaking Mode is available ONLY in invasive ventilation mode.
Patient Preparation
Using Speaking Mode on a patient requires preparation before activating the
mode. Refer to the Sequence for Use diagram in Figure 13-58 and follow the
necessary steps before starting Speaking Mode and discontinuing Speaking
Mode. The Speaking Mode Clinical Guide describes patient care steps in
detail.
WARNING:
Read and understand these Speaking Mode Option instructions before using
Speaking Mode on a patient.
WARNING:
Wait at least 72 hours after a tracheostomy before initiating Speaking
Mode.