Volume control ventilation settings report (vcvs), Rs-232 communications protocol – Philips V200 User Manual
Page 297
Appendix A
RS-232 Communications Protocol
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
Appendix A-11
Volume Control
Ventilation Settings
Report (VCVS)
When the ventilator receives VCVS followed by a carriage return, it will respond
by transmitting the information shown in Table A-2: “Volume Control
Ventilation Settings Report”. The ventilator responds to the VCVS command by
returning a string with a variable length. Fields 2 through 4 define the length
of the message. The last character transmitted is a stop code indicating the
end of the message. The second field indicates the number of characters
between the start and stop codes. The third field indicates the number of
Low EPAP Alarm
Status
NORMAL
N/A
NORMAL
ALARM
◆
RESET
◆
N/A
High Leak Alarm
Status
NORMAL
N/A
NORMAL
ALARM
◆
RESET
◆
N/A
100% O
2
Function
Status
OFF
◆◆◆
N/A
ON
◆◆◆◆
OFF
◆◆◆
N/A
Status of the
100% O
2
LED on
the front panel
Alarm Silence Status
OFF
◆◆◆
N/A
ON
◆◆◆◆
OFF
◆◆◆
N/A
Status of the Alarm
Silence LED on the
front panel
Screen Lock Status
OFF
◆◆◆
N/A
ON
◆◆◆◆
OFF
◆◆◆
N/A
Status of the
Screen Lock LED
on the front panel
Tube Type
ET
◆◆◆◆
N/A
ET
◆◆◆◆
TT
◆◆◆◆
NONE
◆◆
N/A
Tube type as
selected in
Respiratory
Mechanics Screens
(Not used)
Tube Size
7.0
◆◆◆
0.1 for 3.5
to 9.9
1 for 10
3.5 - 10
mm
Tube size as
selected in
Respiratory
Mechanics Screens
(Not used)
VCV High Minute
Volume Alarm Limit
Setting
60.0
◆◆
0.01 for
0.00 to 9.99
0.1 for 10.0
to 60.0
0.00 - 60.0
L
Static C & R Plateau
End Pressure
8.2
◆◆◆
0.1 for -20
to 99.9
1 for 100 to
130
-20.0 - 130 cmH
2
O
P
plat
display from
Static C & R
Screen
Stop Code
0x03
N/A
N/A
N/A
ASCII End
Transmission
Character (ETX)
Ventilation Report (Continued)
Description
Example
Resolution
Range
Units
Comments
Table A-1: Ventilation Report (Sheet 10 of 10)